Model 2701 User’s Manual
Scanning
7-25
2.
Press the or key to display IMM SCAN: N and press ENTER.
a.
Press the or key to enable or disable low limit 1 (LLIM1 SCAN:N/Y) and
press ENTER.
b.
Press the or key to enable or disable high limit 1 (HLIM1 SCAN:N/Y) and
press ENTER.
c.
Press the or key to enable or disable low limit 2 (LLIM2 SCAN:N/Y) and
press ENTER.
d.
Press the or key to enable or disable high limit 2 (HLIM2 SCAN:N/Y) and
press ENTER.
3.
Finish configuring the scan by performing Step 4 and Step 5 of the “Advanced scan
setup procedure,” page 7-16.
NOTE
For a remote programmed monitor scan, use the ROUTe:MONitor:POINts
command to specify the number of channels to scan (Table 7-1).
4.
While in the normal measurement state, select and enable the monitor channel as
explained in “Scan configuration — Monitor channel,” page 7-18.
When the reading limit for the monitor channel is reached, the scan will be
triggered to start. When the monitor channel is scanned, the display will show the
reading that triggered the scan.
If the reading limit event is still present on the monitor channel when the scan
finishes, the scan will be triggered to run again. Note that the scan can also be run
by pressing STEP or SCAN.
5.
To disable the monitor scan, perform the following steps:
a.
To disable monitor, press SHIFT and then MONITOR (MON annunciator
turns off).
b.
To disable limits, press SHIFT and then ON/OFF. Press or to display
“LIMITS: OFF” and press ENTER.
2701-900-01.book Page 25 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...