Model 2701 User’s Manual
Scanning
7-29
Reference
a.
ROUTe:SCAN <clist> — Channels will be scanned in the order that they are
listed. The following example shows the proper format for specifying channels in a
scan list for a sequential scan:
ROUT:SCAN (@101:110,201,204,206)
For the above scan list, the scan will run starting with the lowest numbered channel
(101) and then sequence up (forward) to the highest numbered channel (206).
Remote programming can also be used to run non-sequential scans. Any scan list
configured to scan backward is considered a non-sequential scan. The following
examples configure non-sequential scan lists:
Example 1) ROUT:SCAN (@101:105,103,106:110)
Example 2) ROUT:SCAN (@110:101)
Example 1 — The scan starts with channel 101 and proceeds to channel 105. After
channel 105 is scanned, the unit backs up to scan channel 103, then proceeds
forward to scan channels 106 through 110.
Example 2 — The scan starts with channel 110 then proceeds backward to
channel 101.
Commands
Description
Default Ref
Buffer commands
TRACe:DATA?
Read buffer readings.
TRACe:CLEar
Clear buffer.
Channel list parameter:
<clist> = (@SCH)
where: S =
Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
Notes:
1. Not affected by *RST and SYSTem:PRESet. Front panel factory default is OFF.
2. *RST sets count to 1, and SYSTem:PRESet sets count to INFinity.
3. The default value depends on which switching module is installed.
Table 7-1 (continued)
Scanning commands
2701-900-01.book Page 29 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...