7-34
Scanning
Model 2701 User’s Manual
Table 7-2
External trigger scan example
Front panel operation
Remote programming
1
Restore defaults:
Restore defaults (SHIFT SETUP > RESTORE: FACT).
*RST
2
For front panel operation, proceed to step 3.
For remote programming, clear buffer and disable
buffer auto clear:
TRAC:CLE
TRAC:CLE:AUTO OFF
3
Configure advanced scan:
(SHIFT CONFIG > ADVANCED):
a
Channel 101:
Select TEMP function.
FUNC 'TEMP',(@101)
Configure temperature (SHIFT SENSOR):
Select thermocouple sensor (SENS: TCOUPLE).
TEMP:TRAN TC,(@101)
Select type K thermocouple (TYPE: K).
TEMP:TC:TYPE K,(@101)
Select internal reference junction (JUNC: INT).
TEMP:RJUN:RSEL INT,(@101)
b
Channel 102:
Select
Ω
2 function.
FUNC 'RES',(@102)
Select 1M
Ω
range.
RES:RANG 1e6,(@102)
c
Disable (off) channels 103 through 122 (SHIFT CH-OFF).
ROUT:SCAN (@101,102)
d
Enable immediate scan (IMM SCAN: Y).
ROUT:SCAN:TSO IMM
e
Disable timer (TIMER? NO).
f
Set reading count to infinity (RDG CNT: INF).
TRIG:COUN INF
SAMP:COUN 2
4
Select external trigger control source:
Press EX TRIG.
TRIG:SOUR EXT
5
Enable scan:
Press SCAN.
ROUT:SCAN:LSEL INT
INIT
2701-900-01.book Page 34 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...