8-10
Triggering
Model 2701 User’s Manual
Voltmeter complete
The VMC output provides a TTL-compatible output pulse that can be used to trigger other
instruments. The specifications for this trigger pulse are shown in Figure 8-5. Typically,
you would want the Model 2701 to output a trigger after the settling time of each
measurement.
Figure 8-5
Trigger link output pulse specifications (VMC)
Meter
Complete
TTL High
(3.4V Typical)
TTL Low
(0.25V Typical)
10
μ
s
Minimum
2701-900-01.book Page 10 Wednesday, August 3, 2011 9:43 AM