External triggering example
For a test system that requires a large number of switching channels, the Model 2701 can
be used with external scanners such as the Keithley Models 7001 and 7002. For example,
10 Model 7011s installed in the Model 7002 can provide up to 400 2-pole channels, as
shown in Figure 8-6.
Figure 8-6
DUT test system
The Trigger Link connections for this test system are shown in Figure 8-7. Trigger Link of
the Model 2701 is connected to Trigger Link (either IN or OUT) of the Model 7002. Note
that with the default trigger settings on the Model 7002, line #1 is an input and line #2 is
an output. This complements the trigger lines on the Model 2701.
For this example, the Models 2701 and 7002 are configured as follows:
Model 2701
Factory defaults restored (accessed from SHIFT-SETUP)
External triggers (accessed from EX TRIG)
Buffer enabled and set to store 400 readings
DUT
#1
DUT
#400
DUT
#2
OUTPUT
Card 1
1
2
400
10 7011 MUX Cards
Model 2701
!
2701 ETHERNET MULTIMETER / DATA ACQUISISTION SYSTEM
RANGE
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE
RECALL
OPEN
DCV
DCI
MATH
O U T P U T
RATIO
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
CH AVG
CONT
PERIOD SENSOR
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TYPE
LSYNC
TEST
MONITOR
STEP
SCAN
OCOMP
CH-OFF
CARD
CLOSE
INTEGRA SERIES
RS-232
ETHERNET
CAT I
2701-900-01.book Page 11 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...