Model 2701 User’s Manual
Triggering
8-13
1.
Press EX TRIG to place the Model 2701 in the external trigger mode.
2.
Press STEP on the Model 7002 to take it out of idle and start the scan. The
scanner’s output pulse triggers the Model 2701 to take a reading, store it, and send
a trigger pulse. The following explanation on operation is referenced to the
operation model shown in Figure 8-8.
Figure 8-8
Operation model for triggering example
A.
Pressing EX TRIG on the Model 2701 places it at point A in the flowchart, where it
is waiting for an external trigger.
B.
Pressing STEP on the Model 7002 takes it out of the idle state and places operation
at point B in the flow chart.
C.
For the first pass through the model, the scanner does not wait at point B for a
trigger. Instead, it closes the first channel.
Wait for
Trigger Link
Trigger
Wait for
Trigger Link
Trigger
Scan
Channel
Output
Trigger
Scanned
400
Channels
?
Make
Measurement
Output
Trigger
Trigger
Bypass
Idle
No
Yes
Trigger
Model 7002
Press STEP to start scan
Model 2701
Press EX TRIG
A
B
C
D
E
F
2
1
2701-900-01.book Page 13 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...