Model 2701 User’s Manual
Limits and Digital I/O
9-17
Digital outputs
With the digital outputs of the Model 2701 enabled, the digital outputs will respond as
follows for each resistor reading:
LO limit 2
LO limit 1
HI limit 1
HI limit 2
Resistor
tolerance
bin
Affected
outputs*
Pass
Pass
Pass
Pass
1%
1
None
Pass
Fail
Pass
Pass
5%
2
#1 and #5
Pass
Pass
Fail
Pass
5%
2
#2 and #5
Fail
Fail
Pass
Pass
>5%
3
#1, #3, and #5
Pass
Pass
Fail
Fail
>5%
3
#2, #4, and #5
*Affected outputs are pulled (or pulsed) high or low when a limit test fails.
2701-900-01.book Page 17 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...