Model 2701 User’s Manual
Remote Operations
10-5
•
Enable protected commands. This takes the Model 2701 out of the password-
protected mode. Example using “DEFAULT” as the password:
SYSTem:PASSword:CENable
“DEFAULT”
•
Query the password protection state (enabled or disabled). Example:
SYSTem:PASSword:CENable:STATe?
•
Change the password. The password is case sensitive and can include non-ASCII
characters from 1 to 255. Make sure to enclose the password in quotes. Example to
change password to “INTEGRA”:
SYSTem:PASSword:NEW
“INTEGRA”.
NOTE
The password can be changed from the Model 2701’s internal web page
(see “Internal web page,” page 10-16).
NOTES:
1.
Sending a password-protected command causes error -203 (command protected).
2.
Using an incorrect password in a password command causes error -221 (settings
conflict).
3.
The present password and command protection state are not affected by *RST or
SYSTem:PRESet.
Battery
System settings, such as remote programming interface settings, are saved in non-volatile
memory. The rechargeable lithium ion battery is used for the 450,00 reading buffer. The
system commands to acquire status information about the battery are listed in Table 10-1.
Error codes associated with the battery:
+555 Cannot charge battery
– Will occur approximately one hour after power-up if the
battery cannot charge and needs to be replaced.
+556 Battery removed
– Will occur one time on power-up. Subsequent power-ups will
not repeat this error.
NOTE
Details on the rechargeable lithium ion battery are provided in Section 1 (see
“Battery,” page 1-3).
2701-900-01.book Page 5 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...