Model 2701 User’s Manual
Status Structure
11-17
Figure 11-7
Questionable event status
Condition registers
As Figure 11-1 shows, each status register set (except the Standard Event Register set) has
a condition register. A condition register is a real-time, read-only register that constantly
updates to reflect the present operating conditions of the instrument. For example, while
the Model 2701 is in the idle state, bit B10 (Idle) of the Operation Condition Register will
be set. When the instrument is taken out of idle, bit B10 clears.
The commands to read the condition registers are listed in Table 11-4. For details on read-
ing registers, see “Reading registers,” page 11-7.
Table 11-4
Condition register commands
Command
Description
STATus:OPERation:CONDition?
Read Operation Condition Register.
STATus:MEASurement:CONDition?
Read Measurement Condition Register.
STATus:QUEStionable:CONDition?
Read Questionable Condition Register.
Note: The format of the response messages for the above queries is set by the FORMatt:SREGister
command.
(B13 - B9)
(B14)
(B8)
(B15)
OR
Questionable
Condition
Register
Questionable
Event Enable
Register
Warn = Command Warning
Cal = Calibration Summary
Temp = Temperature Summary
& = Logical AND
OR = Logical OR
0
(B13 - B9)
(B14)
(B8)
(B15)
0
Warn
Cal
Questionable
Event
Register
(B13 - B9)
(B14)
(B8)
(B15)
To
Questionable
Summary Bit
(QSB) of Status
Byte Register
(B7 - B5)
(B7 - B5)
(B7 - B5)
Warn
Cal
0
Warn
Cal
(B4)
Temp
(B4)
Temp
(B3 - B0)
(B3 - B0)
(B3 - B0)
(B4)
Temp
&
&
&
&
2701-900-01.book Page 17 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...