15-18
SCPI Reference Tables
Model 2701 User’s Manual
[SENSe[1]]
:CONTinuity
Path to configure continuity test:
:THReshold <NRf>
Set threshold resistance in ohms
(1 to 1000).
10
:THReshold?
Query threshold resistance.
Notes:
1.
CAVerage:DELay and RATio:DELay are coupled. Changing the delay for channel average also changes the delay for channel
ratio and vice versa.
2.
For 60Hz line power, the default for aperture is 83.33msec. For 50Hz, the default is 100msec.
3.
REPeat is the *RST default and MOVing is the SYSTem:PRESet default. From the front panel, the factory default is MOVing.
4.
OFF is the *RST default and ON is the SYTem:PRESet default.
5.
The following commands can instead be used to select the reference junction and set the simulated reference temperature:
TEMPerature:RJUNction:RSELect <name> [, <clist>]
Select reference junction; SIMulated, INTernal, or EXTernal.
TEMPerature:RJUNction:RSELect? [<clist>]
Query reference junction.
TEMPerature:RJUNction:SIMulated <n> [, <clist>]
Set simulated reference temperature; 0 to 50 (°C), 32 to 122 (°F),
or 273 to 323 (K).
TEMPerature:RJUNction:SIMulated? [<clist>]
Query simulated reference temperature.
6.
When using multiple channel operation (ROUT:MULT command) to connect a switching module channel to the DMM for
thermocouple temperature measurements, the SIMulated reference junction will be used if the INTernal or EXTernal reference
junction is selected.
7.
With a Model 7700, 7706, or 7708 installed, the default sensor junction is Internal. Otherwise, the Simulated (23°C) junction is
selected.
8.
Only one USER RTD per scan list.
Table 15-5 (continued)
SENSe command summary
Command
Description
Default
parameter
Ref
SCPI
2701-900-01.book Page 18 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...