Model 2701 User’s Manual
SCPI Reference Tables
15-27
Table 15-10
UNIT command summary
Command
Description
Default
parameter
Ref
SCPI
UNIT
:TEMPerature <name>
Select temperature units (C, CEL, F,
FAR, or K).
C
:TEMPerature?
Query temperature units.
:VOLTage
Path to configure voltage units.
[:DC] <name>
[, <clist>]
Select DCV measurement units (V or DB).
V
:DB
Path to set DB reference voltage:
:REFerence <n>
Specify reference in volts (1e-7 to 1000).
1
:REFerence?
Query reference.
[:DC]? [<clist>]
Query DCV units.
:AC <name> [, <clist>]
Select ACV measurement units (V or DB).
V
:DB
Path to set DB reference voltage.
:REFerence <n>
Specify reference in volts (le-7 to 1000).
1
:REFerence?
Query DB reference.
:AC? [<clist>]
Query ACV units.
2701-900-01.book Page 27 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...