Model 2701 User’s Manual
Signal Processing Sequence and Data Flow
D-3
Signal processing using instrument features
Figure D-2 shows the processing sequence for an input signal with various instrument
features enabled. If a feature is not enabled, the reading simply falls through to the next
enabled feature or to the display.
Figure D-2
Signal processing using instrument features
Input Signal
Rel
Math
(mX+b, Percent,
or Reciprocal)
Rel’ed reading = Reading - Rel value
Y = mX + b
Percent = ————––—–
X - Reference
Reference
100%
Reciprocal = 1/ X
Limits
Filter
Buffer
Display
Reading
Recall
OComp OComp reading =
Δ
V/
Δ
I
A/D
Conversion
Process
Store
Signal
Conditioning
Output trigger pulse (VMC)
Moving or Repeating
2701-900-01.book Page 3 Wednesday, August 3, 2011 9:43 AM