2-28
Close/Open Switching Module Channels
Model 2701 User’s Manual
Figure 2-10
Testing DUT 1
Test procedure:
NOTES
The following test procedure assumes a Model 7700 switching module installed
in slot 1 of the mainframe.
The procedure assumes that the instrument is operating in the continuous
measurement (triggering) mode (see “Defaults and user setups,” page 1-21).
Do not use the following procedure to perform thermocouple temperature
measurements with the INTernal or EXTernal reference junction selected. The
SIMulated reference junction will instead be used resulting in invalid readings.
The “ERR” annunciator will turn on to indicate that the integrity of the
temperature reading is questionable.
DMM
Model 7700 Switching Module
Model 2701
Slot 1
Equivalent Circuit
DUT
1
Ch 23
(Closed)
HI
LO
HI
LO
HI
LO
HI
LO
HI
LO
DMM
Mutliple channel operation:
Open channels
Close channel 123
Close channel 125
Close channel 101
Close channel 111
DUT
1
External
Source
External
Source
Input
Ch 25
Ch 1
Sense
Ch 24
Ch 11
Sense
2701-900-01.book Page 28 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...