3-34
Basic DMM Operation
Model 2750 Multimeter/Switch System User’s Manual
Temperature measurement configuration
The Model 2750 is configured to measure temperature from the temperature measurement
configuration menu. Use the following general rules to navigate through the menu
structure:
•
Press SHIFT and then SENSOR to enter the menu structure.
•
Cursor position is indicated by a flashing menu item or parameter. Cursor position
is controlled by the
and
keys.
•
With the cursor on a menu item or parameter, use the
Δ
and
∇
keys to scroll
through the available options.
•
A displayed menu item and parameter is selected by pressing ENTER.
•
You can exit from the menu structure by pressing EXIT. However, any ENTERed
selections will apply.
Thermocouple temperature measurement configuration
The steps to configure thermocouple measurements are provided in
. After press-
ing SHIFT and then SENSOR, the menu starts at step 1 to select measurement units.
Each time you press ENTER to make a selection, the menu will automatically go to the
next selection. After pressing ENTER for the last step, the instrument will return to the
normal measurement state.
NOTE
An INT card is a switching module that has an internal reference junction (i.e.,
Model 7700). The INT reference junction setting cannot be selected if there is
not at least one INT card installed in the unit. With no INT cards installed,
selecting INT will cause the “NO INT CARDS” message to be displayed briefly.
With at least one INT card installed, the INT reference junction can be selected.
However, if you select it for the front panel inputs, or for a switching module that
does not have an internal reference junction (i.e., Model 7702), the simulated
(SIM) reference junction will instead be used and the “ERR” annunciator will
turn on.
Table 3-4
Thermocouple temperature measurement configuration
Step
Menu structure
Description
1
UNITS: C, F, or K
Select temperature measurement units (°C, °F, or K).
2
SENS: TCOUPLE
Select the thermocouple transducer.
3
TYPE: J, K, T, E, R, S, B, or N Select thermocouple type.
4
JUNC: SIM, INT, or EXT
Select the SIMulated, INTernal, or EXTernal reference junction*.
SIM: 000°C to 065°C,
273K to 338K, or
032°F to 149°F
For the SIMulated reference junction, set the reference junction
temperature. The displayed units depend on the present UNITS
setting.
5
OPEN DET: Y or N
Enable (Y) or disable (N) the open thermocouple detector.
*When using multiple channel operation (ROUT:MULT command) to connect a switching module input channel to the DMM, the
SIMulated reference junction will be used if the INTernal or EXTernal reference junction is selected.
2750-900-01.book Page 34 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...