Model 2750 Multimeter/Switch System User’s Manual
Basic DMM Operation
3-35
Thermistor temperature measurement configuration
The steps to configure thermistor measurements are provided in
SHIFT and then SENSOR, the menu starts at step 1 to select measurement units.
Each time you press ENTER to make a selection, the menu will automatically go to the
next selection. After pressing ENTER for the last step, the instrument will return to the
normal measurement state.
4-wire RTD temperature measurement configuration
The Alpha, Beta, Delta, and
Ω
at 0°C parameters for the five basic RTD types are provided
. Note that these parameters can be modified using remote programming.
Table 3-5
Thermistor temperature measurement configuration
Step
Menu structure
Description
1
UNITS: C, F, or K
Select temperature measurement units (°C, °F, or K).
2
SENS: THRMSTR
Select the thermistor transducer.
3
TYPE: 2200
Ω
, 5000
Ω
, or 10k
Ω
Select thermistor resistance.
Table 3-6
RTD parameters
Type
Standard
Alpha
Beta
Delta
Ω
at 0°C
PT100
ITS-90
0.00385055
0.10863
1.49990
100
Ω
D100
ITS-90
0.003920
0.10630
1.49710
100
Ω
F100
ITS-90
0.003900
0.11000
1.49589
100
Ω
PT385
IPTS-68
0.003850
0.11100
1.50700
100
Ω
PT3916
IPTS-68
0.003916
0.11600
1.50594
100
Ω
2750-900-01.book Page 35 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...