Channel setup
The <clist> parameter is used to set up scan channels. For example, the following exam-
ples show how to set up scan channel 101:
FUNC 'VOLT', (@101)
' Set 101 for DCV.
VOLT:RANG 10, (@101)
' Set 101 for 10V range.
VOLT:DIG 4.5, (@101)
' Set 101 for 4H digit resolution.
VOLT:NPLC 3, (@101)
' Set 101 rate for 3 PLC.
NOTE
In the above command sequence, channel 101 is first set for DCV before sending
the other commands to set range, digits and rate. If channel 101 was instead set
to a different function (i.e., RESistance), the VOLT commands to set range, dig-
its and rate would generate error +700 (Invalid function in scanlist).
Buffer
For front panel scanning, the reading count specifies the number of readings to store in the
buffer. For remote scanning, the sample count specifies the number of readings to store in
the buffer.
Readings stored in the buffer by the TRAC command (or by front panel data store opera-
tion) must be cleared before sending INITiate or READ? to take the instrument out of idle.
The following command clears the buffer:
TRACe:CLEar
' Clear buffer.
Scanning commands
Scanning commands are listed in
. Additional information on these commands
follow the table.
NOTE
Query commands and optional command words are not included in
The unabridged SCPI tables are provided in
.
2750-900-01.book Page 27 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...