8-16
Triggering
Model 2750 Multimeter/Switch System User’s Manual
The trigger model can be configured to output a trigger after the completion of a series of
measurements, or after every measurement. For example, with the sample counter set to
10 and the trigger counter set to one, a trigger will be sent after the 10 measurements are
performed. If instead, the trigger counter is set to 10 and the sample counter is set to 1, a
trigger will be sent after each measurement.
Triggering commands
Commands for triggering are summarized in
.
Table 8-2
SCPI commands — triggering
Commands
Description
Default Ref
ABORt
Reset trigger system.
a
INITiate[:IMMediate]
Initiate one trigger cycle.
b
INITiate:CONTinuous <b>
Enable/disable continuous initiation; <b> = ON or
OFF.
Note 1
c
FETCh?
Request the last reading(s).
d
READ?
Perform an ABORt, INITiate and a FETCh?
d
TRIGger:SOURce <name>
Select control source; IMMediate, TIMer, MANual,
BUS, or EXTernal.
IMM
e
TRIGger:TIMer <n>
Set timer interval; 0 to 999999.999 (sec).
0.1
TRIGger:COUNt <NRf>
Set trigger count; 1 to 110000 or INFinity.
Note 2
TRIGger:DELay <n>
Set delay; 0 to 999999.999 (sec).
0
TRIGger:DELay:AUTO <b>
Enable or disable auto delay.
f
TRIGger:SIGNal
Loop around control source.
g
SAMPle :COUNt <NRf>
Set sample count; 1 to 110000.
1
h
*RST
Restore *RST defaults (see “Default” column of this
table). Places 2750 in the idle state.
Notes:
1. Defaults for continuous initiation:
SYSTem:PRESet enables continuous initiation.
*RST disables continuous initiation.
2. Defaults for trigger count:
SYSTem:PRESet sets the count to INF (infinity).
*RST sets the count to 1.
2750-900-01.book Page 16 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...