Model 2750 Multimeter/Switch System User’s Manual
Remote Operations
10-3
Separate function setups
A few settings from the front panel are global. That is, the setting on one function also
applies to the other functions. For example, if you set DCV for 3
H
digits, all the other func-
tions will also be set to 3
H
digits. Using remote programming, each function can have its
own unique setup. For example, DCV can be set to 3
H
digits, ACI can be set to 4
H
digits and
DCI can be set to 5
H
digits.
Global settings from the front panel that can be set separately using remote programming
include digits, rate, and filter configuration (except on/off, which can be set separately).
NOTE
Do not confuse function setups with scan channel setups. For scan channels,
separate settings for digits, rate, and filter configuration can be set from either
the front panel or remote programming. See
for details on scan chan-
nel setup.
DCV input divider
Using remote programming, you can enable the DCV input divider for the 100mV, 1V, and
10V ranges. When enabled, the input resistance for these DCV ranges are reduced to
10M
Ω
. See
for details on the DCV input divider.
Multiple channel operation
For normal system channel operation, when one measurement channel is closed, the previ-
ous measurement channel opens. With the use of the ROUTe:MULTiple commands, you
gain independent control of all switching module channels, including the relays that con-
nect the input signal to the DMM. See
for details.
2750-900-01.book Page 3 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...