11-12
Status Structure
Model 2750 Multimeter/Switch System User’s Manual
•
Bit B1
— Not used.
•
Bit B2, Query Error (QYE)
— Set bit indicates that you attempted to read data
from an empty Output Queue.
•
Bit B3, Battery-backed RAM Error (BRE)
— Set bit indicates that the scan can-
not be resumed (error +517) or there was a loss of data (error +516).
•
Bit B4, Execution Error (EXE)
— Set bit indicates that the Model 2750 detected
an error while trying to execute a command.
•
Bit B3, Device-Dependent Error (DDE)
— Set bit indicates that an instrument
operation did not execute properly. Some of the errors specific to the Model 2750
that will set this bit include the following:
•
Error +516: Battery backed RAM error — Data stored in RAM has been lost.
Replace the battery if frequent failures occur.
•
Error +517: Cannot resume scan — Due to a card ID change, auto scan has
disabled. The scan will not resume after a power interruption. For details, see
“Scan configuration,” page 7-11
.”
•
Error +520: Saved setup scancard mismatch — Settings for a user setup or
power-on setup do not match the switching module types presently installed.
For details, see
“Defaults and user setups,” page 1-19
.”
•
Error +523: Card hardware error — Communication with the microprocessor
on a switching module card has been lost.
•
Error +524: Unsupported card detected — The Model 2750 has detected an
installed Model 77XX switching module that is not supported by the current
version of firmware.
•
Bit B6, User Request (URQ)
— Set bit indicates that the LOCAL key on the
Model 2750 front panel was pressed.
•
Bit B7, Power On (PON)
— Set bit indicates that the Model 2750 has been turned
off and turned back on since the last time this register has been read.
Operation event register
The bits of the Operation Event Register (
) are described as follows:
•
Bits B0 through B3
— Not used.
•
Bit B4, Measuring (Meas)
— Set bit indicates that the instrument is performing a
measurement.
•
Bit B5, Waiting for Trigger (Trig)
— Set bit indicates that the Model 2750 is in
the trigger layer waiting for a trigger event to occur.
•
Bits B6 and B7
— Not used.
•
Bits B8, Filter Settled (Filt)
— Set bit indicates that the filter has settled or the fil-
ter is disabled.
•
Bit B9
— Not used.
•
Bit B10, Idle State (Idle)
— Set bit indicates the Model 2750 is in the idle state.
•
Bits B11 through B15
— Not used.
2750-900-01.book Page 12 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...