Model 2750 Multimeter/Switch System User’s Manual
Status Structure
11-21
On power-up, all error messages are enabled and will go into the Error Queue as they
occur. Status messages are not enabled and will not go into the queue. As listed in
, there are commands to enable and/or disable messages. For these commands,
the <list> parameter is used to specify which messages to enable or disable. The messages
are specified by their codes. The following examples show various forms for using the
<list> parameter.
<list>
= (-110)
Single message
= (-110:-222)
Range of messages (-110 through -222)
= (-110:-222, -220)
Range entry and single entry (separated by a comma)
When you enable messages, messages not specified in the list are disabled. When you dis-
able messages, each listed message is removed from the enabled list.
NOTE
To prevent all messages from entering the Error Queue, send the enable com-
mand along with the null list parameter as follows:
STATus:QUEue:ENABle ().
Programming example — read error queue
The following command reads the error queue:
STAT:QUE?
Table 11-6
Error queue commands
Command
Description
Notes
STATus:QUEue:[:NEXT]?
Read and clear oldest error/status (code and message).
1
STATus:QUEue:ENABle <list>
Specify error and status messages for Error Queue.
2
STATus:QUEue:ENABle?
Read the enabled messages.
STATus:QUEue:DISable <list>
Specify messages not to be placed in queue.
2
STATus:QUEue:DISable?
Read the disabled messages.
STATus:QUEue:CLEar
Clear messages from Error Queue.
SYSTem:ERRor?
Read Error Queue.
1
SYSTem:CLEar
Clear messages from Error Queue.
Notes:
1. Power-up and *CLS empties the Error Queue. STATus:PRESet has no effect.
2. Power-up enables error messages and disables status messages. *CLS and STATus:PRESet have no effect.
2750-900-01.book Page 21 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...