Model 2750 Multimeter/Switch System User’s Manual
Temperature Equations
F-5
Table F-6
Type R inverse function polynomial
-50°C to 250°C
(-226µV to
1,923µV)
250°C to 1,200°C
(1,923µV to
13,228µV)
1,064°C to 1,664.5°C
(11,361µV to
19,739µV)
1,664.5°C to 1,768.1°C
(19,739µV to
21,103µV)
c
0
=
0.0
1.334 584 505
×
10
1
-8.199 599 416
×
10
1
3.406 177 836
×
10
4
c
1
=
1.889 138 0
×
10
-1
1.472 644 573
×
10
-1
1.553 962 042
×
10
-1
-7.023 729 171
c
2
=
-9.383 529 0
×
10
-5
-1.844 024 844
×
10
-5
-8.342 197 663
×
10
-6
5.582 903 813
×
10
-4
c
3
=
1.306 861 9
×
10
-7
4.031 129 726
×
10
-9
4.279 433 549
×
10
-10
-1.952 394 635
×
10
-8
c
4
=
-2.270 358 0
×
10
-10
-6.249 428 360
×
10
-13
-1.191 577 910
×
10
-14
2.560 740 231
×
10
-13
c
5
=
3.514 565 9
×
10
-13
6.468 412 046
×
10
-17
1.492 290 091
×
10
-19
c
6
=
-3.895 390 0
×
10
-16
-4.458 750 426
×
10
-21
c
7
=
2.823 947 1
×
10
-19
1.994 710 149
×
10
-25
c
8
=
-1.260 728 1
×
10
-22
-5.313 401 790
×
10
-30
c
9
=
3.135 361 1
×
10
-26
6.481 976 217
×
10
-35
c
10
=
-3.318 776 9
×
10
-30
Error: 0.02°C to -0.02°C
0.005°C to -0.005°C
0.001°C to -0.0005°C
0.002°C to -0.001°C
t
90
= c
0
+ c
1
E + c
2
E
2
+ c
3
E
3
... c
i
E
i
where: t
90
is the calculated temperature in °C.
E is the measured voltage in microvolts.
Table F-7
Type S inverse function polynomial
-50°C to 250°C
(-235µV to 1,874µV)
250°C to 1,200°C
(1,874µV to 11,950µV)
1,064°C to 1,664.5°C
(10,332µV to 17,536µV)
1,664.5°C to 1,768.1°C
(17,536µV to 18,693µV)
c
0
=
0.0
1.291 509 199
×
10
1
-8.087 801 117
×
10
1
5.333 875 126
×
10
4
c
1
=
1.849 494 60
×
10
-1
1.466 298 863
×
10
-1
1.621 573 104
×
10
-1
-1.235 892 298
×
10
1
c
2
=
-8.005 050 62
×
10
-5
-1.534 713 402
×
10
-5
-8.536 869 453
×
10
-6
1.092 657 613
×
10
-3
c
3
=
1.022 374 30
×
10
-7
3.145 945 973
×
10
-9
4.719 686 976
×
10
-10
-4.265 693 686
×
10
-8
c
4
=
-1.522 485 92
×
10
-10
-4.163 257 839
×
10
-13
-1.441 693 666
×
10
-14
6.247 205 420
×
10
-13
c
5
=
1.888 213 43
×
10
-13
3.187 963 771
×
10
-17
2.081 618 890
×
10
-19
c
6
=
-1.590 859 41
×
10
-16
-1.291 637 500
×
10
-21
c
7
=
8.230 278 80
×
10
-20
2.183 475 087
×
10
-26
c
8
=
-2.341 819 44
×
10
-23
-1.447 379 511
×
10
-31
c
9
=
2.797 862 60
×
10
-27
8.211 272 125
×
10
-36
Error: 0.02°C to -0.02°C
0.01°C to -0.01°C
0.0002°C to -0.0002°C 0.002°C to -0.002°C
t
90
= c
0
+ c
1
E + c
2
E
2
+ c
3
E
3
... c
i
E
i
where: t
90
is the calculated temperature in °C.
E is the measured voltage in microvolts.
2750-900-01.book Page 5 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...