Close/Open Switching Module Channels
2-3
Switching module installation and connections
In order to exercise close/open operations explained in this section, a switching module (or
pseudocard) must be installed in the mainframe. A switching module can be installed by
the user, however external connections to the switching module are only to be performed
by qualified service personnel.
NOTE
For inexperienced users, it is recommended that DUT and external circuitry not
be connected to switching modules. This will allow you to exercise close/open
operations without the dangers associated with live test circuits.
WARNING
To prevent electric shock that could result in injury or death, NEVER
handle a switching module that has power applied to it:
•
Before installing (or removing) a switching module, make sure the
Model 2750 is turned off and disconnected from line power.
•
If the switching module is already connected to DUT, make sure
power is removed from all external circuitry.
Module installation
WARNING
Slot covers must be installed on unused slots to prevent personal con-
tact with high voltage circuits.
Perform the following steps to install a switching module into the Model 2750 mainframe:
1.
Turn the Model 2750 off and disconnect the power line cord and any other cable
connected to the rear panel.
2.
Position the Model 2750 so you are facing the rear panel.
3.
Remove the slot cover plate from the desired mainframe slot. Retain the plate and
screws for future use.
4.
With the top cover of the switching module facing up, slide the module into an
empty slot. For the last
G
inch or so, press in firmly to mate the module connector to
the mainframe connector.
5.
On each side of the module, there is a mounting screw. Tighten these two screws to
secure the module to the mainframe. Do not overtighten.
6.
Reconnect the power line cable and any other cables to the rear panel.
7.
When you turn on the Model 2750, the model number of the switching module will
be briefly displayed.
2750-900-01.book Page 3 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...