2-6
Close/Open Switching Module Channels
Model 2750 Multimeter/Switch System User’s Manual
System channel operation
The system channel is a closed measurement channel that is internally connected to the
internal DMM Input of the Model 2750. The system channel number is displayed on the
Model 2750. For a 4-wire function (i.e.,
Ω
4), the paired channel for the system channel is
internally connected to DMM Sense. The paired channel is not displayed on the Model
2750. When triggered, the DMM performs a measurement and displays it on the Model
2750.
The system channel is selected by closing a measurement channel using the system chan-
nel close keys. These include the
and
keys, or the CLOSE key (SINGLE menu
option). See
“Controlling the system channel,” page 2-9
, for details.
Other important points about system channel operation include the following:
•
There can only be one system channel. This is the channel that is presently dis-
played (and closed) on the Model 2750. When a channel is not displayed, there is
no system channel.
•
When a measurement channel is closed, the input backplane isolation channel also
closes to connect the system channel to DMM input. For a 4-wire function, the
paired channel and the sense backplane isolation channel also close to make the
sense connections to the DMM.
•
When a different measurement channel is closed, the previous system channel
opens. The newly closed (and displayed) measurement channel becomes the sys-
tem channel.
•
The system channel close keys can only close measurement channels that will
automatically connect to the DMM. Non-measurement channels cannot be closed
by the system channel close keys.
NOTE
Use the VIEW option of the CARD menu to display all closed channels in the
mainframe (see
).
2750-900-01.book Page 6 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...