Model 2750 Multimeter/Switch System User’s Manual
Close/Open Switching Module Channels
2-11
Remote programming — system channel control commands
The commands to close and open the system channel are listed in
. When a sys-
tem channel reading is returned, the system channel number will be included in the data
string if the CHANnel data element is selected. The FORMat:ELEMents command is used
to specify the data elements to be included in the data string (see FORMat commands in
Reference:
a.
ROUTe:CLOSe <clist>
This command functions the same as the front panel CLOSE key (SINGLE menu
option) to select the system channel. Only one measurement channel can be
specified in the <clist>.
Trying to close an invalid channel (such as a non-measurement channel) with this
command will result in error -222 (Parameter data out of range).
b.
ROUTe:CLOSe:STATe? <clist>
This query returns a “0” (open) or “1” (closed) for every measurement channel
specified the <clist>. For example, assume <clist> = (@101, 104, 107, 102). The
response message “0, 0, 1, 0” indicates that channel 107 is closed.
The state of non-measurement channels cannot be checked with this command.
Table 2-1
System channel control commands
Commands
Description
Ref
ROUTe:CLOSe <clist>
Specify one measurement channel to close.
a
ROUTe:CLOSe:STATe? <clist> Query closed channels in specified list
(1 = closed).
b
ROUTe:CLOSe?
Returns a <clist> of closed measurement
channels.
c
ROUTe:OPEN:ALL
Open all channels, and disable ratio and channel
average.
d
Channel list parameter:
<clist> = (@SCH)
where: S
= Mainframe slot number (1, 2, 3, 4 or 5)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
2750-900-01.book Page 11 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...