2-20
Close/Open Switching Module Channels
Model 2750 Multimeter/Switch System User’s Manual
Remote programming example (multiple channel operation)
The following example assumes a Model 7700 installed in slot 1. This command sequence
connects channel 101 to channel 111 (through channel 123). Note that these two closed
channels will be internally isolated from the DMM since the backplane isolation channels
(124 and 125) will be open.
ROUT:OPEN:ALL
' Open all channels.
ROUT:MULT:CLOS (@101,111,123)
' Close channels 101, 111 and 123.
When finished with multiple channel operation it is a good, safe practice to open all chan-
nels (
ROUT:OPEN:ALL
).
Multiple channel operation anomalies
•
Anomaly #1
— When you use multiple channel operation to open the system
channel, the channel will open but the system channel number will still be
displayed on the Model 2750. For details, see “
.”
•
Anomaly #2
— For a 4-wire function, you can use multiple channel operation to
open the paired channel. If you then use system channel operation to again select
the already closed system channel, it will not re-close the paired channel. For
details, see “
Anomaly #2 example — opening the paired channel
NOTE
The following anomaly examples assume a Model 7700 installed in slot 1.
Anomaly #1 example — wrong channel displayed
The following example closes channel 102 and connects it to the DMM Input. However,
the Model 2750 will not display the measurement channel that is closed. It will display
channel 101 instead of channel 102.
1.
Use the ALL option for the OPEN key to open all channels in the mainframe.
Remote programming:
ROUT:OPEN:ALL
2.
Press the
key to close (and display) channel 101. This closes channel 101
(which is the system channel) and channel 125 to connect it to the DMM Input
(
).
Remote programming:
ROUT:CLOS (@101)
3.
Use the MULTI option for the CLOSE key to close channel 102. The system chan-
nel is not affected. Channels 101, 102, and 125 are now closed.
Remote programming:
ROUT:MULT:CLOS (@102)
4.
Use the MULTI option for the OPEN key, open channel 101. Even though channel
101 is still being displayed on the Model 2750, it is channel 102 that is actually
connected to the DMM Input (channels 102 and 125 closed).
Remote programming:
ROUT:MULT:OPEN (@101)
2750-900-01.book Page 20 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...