Model 2750 Multimeter/Switch System User’s Manual
Close/Open Switching Module Channels
2-27
Identifying installed modules and
viewing closed channels
On power-up, the model numbers of installed switching modules are displayed briefly.
While in the normal display state, red slot indicators on the right side of the display indi-
cate which slots have a switching module or pseudocard installed.
NOTE
If a Model 7700, 7701, 7702, 7703, 7705, 7708, or 7709 switching module is
removed while the Model 2750 is on, the slot indicator for that slot will remain
on, and the instrument will operate as if the module is installed. That is, the
Model 2750 will operate as if the pseudocard is installed.
NOTE
If a Model 7706 or 7707 is removed while power is on, error +523 (“Card hard-
ware error”) will occur, and the module will be removed from the system.
NOTE
In general, it is not recommended to install or remove switching modules with
the power on.
CARD menu
The CARD menu identifies the switching modules installed in the mainframe, and is used
for the following operations:
•
Configure digital inputs and outputs, and analog outputs for switching modules
that have one or more of those capabilities (i.e., Models 7706 and 7707).
•
View the analog input channels that are presently closed. Also, read digital input
and output ports, and analog output values for switching modules that have one or
more of those capabilities.
Menu navigation keys
— Once in the menu structure, the manual range keys (
Δ
and
∇
)
and the cursor keys (
and
) are used to display menu items and options, and set
parameter values. With the desired item, option or setting displayed, press the ENTER key
to select it. You can cancel a pending selection (and exit the menu structure) by pressing
the EXIT key.
Press the SHIFT key and then the CARD key to display the CARD menu. The Card menu
tree is shown in
. The items and options of the menu are explained as follows:
NOTE
Identifying installed modules — If you simply want to identify installed modules
or pseudocards, select CONFIG or VIEW and use the
Δ
or
∇
key to check each
slot. While in the menu structure, the slot indicator for the selected slot will be
on. The other indicators will be off. When finished, press EXIT.
2750-900-01.book Page 27 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...