3-36
Basic DMM Operation
Model 2750 Multimeter/Switch System User’s Manual
The steps to configure 4-wire RTD measurements are provided in
. After press-
ing SHIFT and then SENSOR, the menu starts at step 1 to select measurement units.
Each time you press ENTER to make a selection, the menu will automatically go to the
next selection. After pressing ENTER for the last step, the instrument will return to the
normal measurement state.
NOTE
As shown in
, you can select the USER sensor type, but you cannot
change the USER parameters from the front panel. The parameters for the
USER type can only be set using remote programming (see TEMPerature:FRTD
commands in
).
Temperature measurement procedure
NOTE
Make sure the INPUTS switch is in the correct position. To use front panel
inputs, it must be in the “F” (out) position. For switching modules, it must be in
the “R” (in) position.
1.
If a switching channel is presently closed (displayed), press OPEN to open it.
2.
Select the temperature measurement function by pressing TEMP.
3.
Configure the temperature measurement as previously explained in “
.”
4.
Connect the temperature transducer(s) to be measured.
5.
If using a switching module, perform the following steps to close the desired chan-
nel. Keep in mind, that for 4-wire RTD measurements, you will close the primary
(INPUT) channel (1 through 10). The channel that it is paired to will close
automatically.
a.
Press the CLOSE key.
b.
Use
,
,
Δ
, and
∇
to key in the channel number and press ENTER. The
previously closed channel(s) (if any) will open, and the specified channel (or
channel pair) will close.
Table 3-7
4-wire RTD temperature measurement configuration
Step
Menu Structure
Description
1
UNITS: C, F, or K
Select temperature measurement units
(°C, °F, or K).
2
SENS: 4W-RTD
Select the 4-wire RTD transducer.
3
TYPE:PT100, D100, F100,
PT385, PT3916, or USER
Select 4-wire RTD type.
2750-900-01.book Page 36 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...