3-48
Basic DMM Operation
Model 2750 Multimeter/Switch System User’s Manual
Basic measurement programming examples
Example #1
—
continuous triggering
The following command sequence places the Model 2750 in a continuous trigger mode to
measure ACV. Whenever DATA? is sent, the last measured reading will be sent to the
computer when the Model 2750 is addressed to talk.
SYST:PRES
' Continuous measurement mode (INIT:CONT ON).
FUNC 'VOLT:AC'
' Select ACV function.
DATA?
' Request last measured reading.
Example #2
—
one-shot triggering
The following command sequence places the Model 2750 in a one-shot trigger mode to
measure offset-compensated ohms. Whenever READ? is sent, a measurement will be trig-
gered, and the measured reading will be sent to the computer when the Model 2750 is
addressed to talk.
*RST
' One-shot measurement mode (INIT:CONT OFF).
FUNC 'FRES'
' Select
Ω
4 function.
FRES:RANG 1e3
' Select 1k
Ω
range.
FRES:OCOM ON
' Enable offset-compensated ohms.
READ?
' Trigger and return one reading.
Example #3
—
temperature measurement using Model 7700
The following command sequence places the Model 2750 in a one-shot trigger mode to
perform a thermocouple temperature measurement at channel 101 (Model 7700 switching
module installed in slot 1). With channel 101 closed, the INIT command triggers one mea-
surement, and the DATA? command sends the measured reading to the computer when it
is addressed to talk.
*RST
' One-shot measurement mode (INIT:CONT OFF).
FUNC 'TEMP'
' Select TEMP function.
UNIT:TEMP F
' Select °F TEMP units.
TEMP:TRAN TC
' Select thermocouple transducer.
TEMP:TC:TYPE J
' Select type J thermocouple.
TEMP:RJUN:RSEL SIM
' Select simulated reference junction.
TEMP:RJUN:SIM 32
' Set reference temperature to 32°F (ice point).
ROUT:CLOS (@101)
' Close channel 101.
INIT
' Trigger one measurement.
DATA?
' Return measured reading.
2750-900-01.book Page 48 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...