4-22
Range, Digits, Rate, Bandwidth, and Filter
Model 2750 Multimeter/Switch System User’s Manual
Filter programming examples
Example #1
— The following command sequence configures filtering for the DCI
function:
CURR:TCON MOV
' Select the moving filter.
CURR:AVER:WIND 0.01
' Set filter window to 0.01%.
CURR:AVER:COUN 10
' Set to filter 10 readings.
CURR:AVER ON
' Enable filter.
Example #2
— The following command sequence configures channels 101 through 115
of the Model 7700 to use the repeat filter when they are scanned.
FUNC 'VOLT'
' Select DCV function.
VOLT:AVER:TCON REP
' Select the repeating filter.
VOLT:AVER:COUN 20,(@101:115)
' Set to filter 20 readings.
VOLT:AVER ON,(@101:115)
' Enable filter.
2750-900-01.book Page 22 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...