Model 2750 Multimeter/Switch System User’s Manual
Rel, Math, Ratio, Channel Average, dB
5-7
value. When rel is set using :ACQuire, the :REFerence? query command returns the
acquired rel value.
Rel programming examples
Example #1
— The following command sequence zeroes the display for DCV.
FUNC 'VOLT'
' Select DCV.
VOLT:REF:ACQ
' Use input level as rel value for DCV.
VOLT:REF:STAT ON
' Enable rel.
Example #2
— The following command sequence configures channel 101 of the
Model 7700 to enable rel and use a 1V rel value when it is scanned.
FUNC 'VOLT',(@101)
' Select DCV for channel 101.
VOLT:REF 1,(@101)
' Set 1V rel value.
VOLT:REF:STAT ON,(@101)
' Enable rel.
Example #3
— The following command sequence configures channel 101 of the
Model 7700 to zero correct the DCV input when it is scanned.
FUNC 'VOLT',(@101)
' Select DCV for channel 101.
ROUT:CLOS (@101)
' Close channel 101.
VOLT:REF:ACQ (@101)
' Use input to channel 101 as rel value.
VOLT:REF:STAT ON,(@101)
' Enable rel.
2750-900-01.book Page 7 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...