Model 2750 Multimeter/Switch System User’s Manual
Scanning
7-11
Scan configuration
A scan is configured from the scan configuration menu which is accessed by pressing
SHIFT and then CONFIG.
shows the basic flowchart to configure a scan. After
entering the menu structure you can configure a simple scan, an advanced scan, or reset
the configuration to the default setup for a simple scan. Refer to the flowchart in
for the following discussions on “
,” “
,” and “
NOTE
Only a sequential scan can be configured from the front panel. For a sequential
scan, the scan proceeds from the lowest numbered channel to the highest.
Non-sequential (backward) scanning is only intended to be performed using
remote programming. Unexpected results may occur if a non-sequential scan is
run from the front panel.
For more information, see “
.
Figure 7-3
Scan configuration flowchart
NOTE
The instrument is always configured to run a scan. On power-up, each available
channel uses the power-on default setup. For example, for factory power-on
default settings, and two Model 7700s installed, the instrument will scan chan-
nels 101 through 220 when the scan is run. See
for details on power-
on default settings.
SHIFT
CONFIG
Simple
Advanced
Reset
Min Chan
Max Chan
Setup
Imm Scan?
Timer?
Rdg Cnt
Timer?
Rdg Cnt
2750-900-01.book Page 11 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...