8-10
Triggering
Model 2750 Multimeter/Switch System User’s Manual
1.
Press EX TRIG to place the Model 2750 in the external trigger mode.
2.
Press STEP on the Model 7002 to take it out of idle and start the scan. The scan-
ner’s output pulse triggers the Model 2750 to take a reading, store it, and send a
trigger pulse. The following explanation on operation is referenced to the operation
model shown in
.
Figure 8-8
Operation model for triggering example
A.
Pressing EX TRIG on the Model 2750 places it at point A in the flowchart, where it
is waiting for an external trigger.
B.
Pressing STEP on the Model 7002 takes it out of the idle state and places operation
at point B in the flow chart.
Wait for
Trigger Link
Trigger
Wait for
Trigger Link
Trigger
Scan
Channel
Output
Trigger
Scanned
400
Channels
?
Make
Measurement
Output
Trigger
Trigger
Bypass
Idle
No
Yes
Trigger
Model 7002
Press STEP to start scan
Model 2750
Press EX TRIG
A
B
C
D
E
F
2
1
2750-900-01.book Page 10 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...