Model 2750 Multimeter/Switch System User’s Manual
Limits and Digital I/O
9-7
Logic sense
The selected logic sense (active high or active low) determines if an output is pulled high
or low when the limit is reached. If logic sense is set high, the output line will be pulled
high when the reading reaches or exceeds the limit. If logic sense is set low, the output line
will be pulled low to 0V when the reading reaches or exceeds the limit.
Pulse option
Pulse option is available for the digital outputs. When enabled, an output line will pulse
high or low (depending on the logic sense setting) for each reading that reaches or exceeds
the limit. The factory default time duration for the pulse is 2ms (maximum), but can be set
from 0.001 to 99999.999 seconds using remote programming. Pulse time cannot be set
from the front panel.
NOTE
The commands to set pulse time and enable/disable pulse output are listed in
. See “Digital output commands” in the table.
The pulse time does not affect measurement speed. If a subsequent in-limit reading occurs
while the output line is being pulsed, the line will be released immediately (pulse
terminated).
Master limit latch
The master limit line is pulled high or low when one or more of the other four limits are
reached or exceeded. The master limit line can be programmed to release when a reading
is inside all four limits, or the master limit can be latched when a failure occurs. When
latched, the master limit line will not release until operation within the trigger model
returns to and passes the control source (see
When scanning, the latched master limit line will not release until the scan is finished and
another scan is started. For example, if after testing a resistor network the master limit line
did set, then the network has passed all tests.
2750-900-01.book Page 7 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...