9-12
Limits and Digital I/O
Model 2750 Multimeter/Switch System User’s Manual
Perform the following steps to enable and configure digital outputs:
1.
Press SHIFT and then OUTPUT.
2.
If the digital output is already on (DOUTPUT: ON), proceed to step 3. Otherwise,
press
to move the cursor to the right, press
Δ
or
∇
key to display “ON,” and
press ENTER.
3.
Use the
∇
key to display the master limit latch (MASTR LATCH) setting: N (no)
or Y (yes).
4.
If you want to retain the present master limit setting, proceed to step 5. Otherwise,
press
to move the cursor to the right, press
Δ
or
∇
key to display “Y” or “N,”
and press ENTER.
5.
Use the
∇
key to display the present logic sense (LSENSE) setting: HIGH or
LOW.
6.
If you want to retain the present logic sense setting, proceed to step 7. Otherwise,
press
to move the cursor to the right, press
Δ
or
∇
key to display “HIGH” or
“LOW,” and press ENTER.
7.
Use the
∇
key to display the present PULSE mode setting: NO or YES.
8.
To retain the present pulse mode setting, press ENTER. Otherwise, press
to
move the cursor to the right, press
Δ
or
∇
key to display “NO” or “YES,” and
press ENTER.
Scanning
While limits can be configured on a per scan channel basis, the digital output configura-
tion cannot. Therefore, for all scan channels that are set to use limits, the digital output
will function according to how the Model 2750 is set up when the scan is run.
2750-900-01.book Page 12 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...