Model 2750 Multimeter/Switch System User’s Manual
SCPI Signal Oriented Commands
13-3
CONFigure:<function> [<rang>], [<res>], [<clist>]
CONFigure:VOLTage[:DC] [<rang>], [<res>], [<clist>]
Configure DCV
CONFigure:VOLTage:AC [<rang>], [<res>], [<clist>]
Configure ACV
CONFigure:CURRent[:DC] [<rang>], [<res>], [<clist>]
Configure DCI
CONFigure:CURRent:AC [<rang>], [<res>], [<clist>]
Configure ACI
CONFigure:RESistance [<rang>], [<res>], [<clist>]
Configure
Ω
2
CONFigure:FRESistance [<rang>], [<res>], [<clist>]
Configure
Ω
4
CONFigure:FREQuency [<rang>], [<res>], [<clist>]
Configure FREQ
CONFigure:PERiod [<rang>], [<res>], [<clist>]
Configure PERIOD
CONFigure:TEMPerature [<clist>]
Configure TEMP
CONFigure:CONTinuity [<clist>]
Configure CONT
Parameters
<rang> = Range parameter for the specified function. For example, for
DCV, range parameter value 10 selects the 10V range. See
the “NOTES” that follow
for additional information.
<res> =
0.1
i.e., 100.0 V (3
H
digits)
0.01
i.e., 10.00 V (3
H
digits)
0.001
i.e., 1.000 V (3
H
digits)
0.0001
i.e., 1.0000 V (4
H
digits)
0.00001
i.e., 1.00000 V (5
H
digits)
0.000001
i.e., 1.000000 V (6
H
digits)
The resolution of the <res> parameter value and the selected range
sets the number of display digits. As shown above, with the 100V
range selected and <res> = 0.1, a 100V reading will be displayed
as 100.0 V (3
H
digits).
The display will default to 3
H
digits when using parameter values
that attempt to set the display below 3
H
digits. For example, a 10V
reading using <res> = 0.1 for the 10V range is displayed as 10.00
V, not 10.0 V.
A command using parameter values that attempt to set the dis-
play above 7
H
digits is ignored, and generates error -221 (set-
tings conflict error).
The <res> parameter is ignored when a <clist> is included in
the command string. Resolution for the scanlist channel(s) is
determined by the present setting for the specified function and
by the present resolution setting for the specified function.
for additional infor-
mation.
<clist> =
Channel(s) in the scanlist to be configured. When the channel
list parameter (<clist>) is included, the present instrument set-
tings are not affected. Instead, the channel(s) in the <clist> for
the specified function is configured. See the “NOTES” that fol-
low
for additional information.
2750-900-01.book Page 3 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...