13-6
SCPI Signal Oriented Commands
Model 2750 Multimeter/Switch System User’s Manual
If the instrument is in the idle state, INITiate takes the instrument out of
the idle state. If continuous initiation is enabled, (INITiate:CONTinuous
ON), then the INITiate command generates error -213 (init ignored).
NOTE
If continuous initiation is enabled, the READ? command will generate
error -213 (init ignored). It can be disabled by sending INITiate:CONTinuous
OFF. The *RST command can also be used to disable continuous initiation. It
also places the Model 2750 in the “one-shot” measurement mode:
*RST
‘ Disable continuous initiation and place 2750 in “one-shot” mode.
READ? ‘ Trigger and return one reading.
NOTE
When readings are stored in the buffer by the TRACe command (or by front
panel data store operation), INIT and multi-sample READ? queries are locked
out. With readings in the buffer that were stored in that manner, you cannot use
the INIT or READ? command if sample count is >1 (error -225, out of memory).
Buffer operation is covered in
.
The buffer of the Model 2750 is nonvolatile. Therefore, readings stored in the
buffer are not lost when the instrument is turned off, or when *RST or
SYSTem:PRESet is sent. When writing test programs that perform multi-sample
measurements (SAMPle:COUNTt >1), you may want to add the TRACe:CLEar
command at the beginning to clear the buffer. However, be careful not to inad-
vertently clear stored readings that are needed.
2750-900-01.book Page 6 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...