Model 2750 Multimeter/Switch System User’s Manual
SCPI Reference Tables
15-3
Table 15-1
CALCulate command summary
Command
Description
Default
parameter
Ref
SCPI
CALCulate[1]
Subsystem to control CALC 1:
✓
:FORMat <name>
[<, clist>]
Select math format (NONE, MXB, PERCent, or
RECiprocal).
PERCent
✓
:FORMat? [<clist>]
Query math format.
✓
:KMATh
Path to configure math calculations:
:MMFactor <NRf>
[, <clist>]
Set “m” factor for mx+b (-4294967295 to
+4294967295).
1
1
:MA1Factor <NRf>
[, <clist>]
Set “m” factor for mx+b (-4294967295 to
+4294967295).
1
1
:MMFactor?
[<clist>]
Query “m” factor.
1
:MA1Factor?
[<clist>]
Query “m” factor.
1
:MBFactor <NRf>
[, <clist>]
Set “b” factor for mx+b (-4294967295 to
+4294967295).
2
0
:MA0Factor <NRf>
[, <clist>]
Set “b” factor for mx+b (-4294967295 to
+4294967295).
2
0
:MBFactor? [<clist>]
Query “b” factor.
2
:MA0Factor?
[<clist>]
Query “b” factor.
2
:MUNits <char>
Specify units for mx+b reading.
3
‘X’
:MUNits?
Query “mx+b” units.
:PERCent <NRf>
[, <clist>]
Set target value for PERCent calculation
(-4294967295 to +4294967295).
1
:ACQuire
Use input signal as target value.
:PERCent? [<clist>]
Query percent.
:STATe <b> [, <clist>]
Enable or disable kmath calculation.
(Note 4)
✓
:STATe? [< clist>]
Query state of kmath function.
✓
:DATA?
Read result of kmath calculation.
✓
CALCulate2
Subsystem to control CALC 2:
✓
:FORMat <name>
Select math format: (MEAN, SDEViation,
MAXimum, MINimum, PKPK, or NONE).
NONE
✓
:FORMat?
Query math format.
✓
:STATe <b>
Enable or disable calculation.
(Note 4)
✓
:STATe?
Query state of math function.
✓
:IMMediate
Recalculate raw input data in buffer.
✓
:IMMediate?
Perform calculation and read result.
✓
:DATA?
Read math result of CALC 2.
✓
2750-900-01.book Page 3 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...