Model 2750 Multimeter/Switch System User’s Manual
Close/Open Switching Module Channels
2-33
Model 7700 switching module
NOTE
Connection and wiring procedures for the Model 7700 are to be performed by
qualified service personnel. This information is provided in
7700 Connection Guide).
Switching module capabilities
Channels 1 through 20
— The Model 7700 can multiplex one of 20 2-pole signals, or one
of 10 4-pole signals into the input of the Model 2750.
Channels 21 and 22
— The Model 7700 can multiplex one of two 2-pole current signals
into the input of the Model 2750.
CAUTION
To prevent damage to the Model 7700 switching module, do not exceed
these maximum signal levels:
Channels 1-20:
300VDC or RMS, 1A switched, 60W, 125VA
Channels 21, 22:
60VDC or 30V RMS, 3A switched, 60W, 125VA
NOTE
System channel operation
– Of the 22 measurement channels, only one channel
(or channel pair) can be closed at the same time. When you close a channel (or
channel pair), all other measurement channels will open. The user has no con-
trol of channels 23, 24, and 25. The open/close state of these channels are deter-
mined by the selected function.
The Model 7700 has six temperature transducers to monitor the cold junction temperature
at the screw terminals. For temperature measurements, this internal reference junction
allows thermocouples to be connected directly to the screw terminals of the module.
When the Model 2750 is on the DCV, ACV,
Ω
2, CONT,
Ω
4, FREQ, PERIOD, or TEMP
function, channels 1 through 20 are available. When on a current function (DCI or ACI),
channels 21 and 22 are the only available channels.
The Model 7700 can accommodate 4-wire measurements by using channel pairs. Primary
channels 1 through 10 become paired to channels 11 through 20. For example, with the
Ω
4
function selected, channel 1 becomes paired to channel 11. For example, when you close
channel 1, channel 11 will also close.
The 2-wire functions include DCV, ACV, DCI, ACI,
Ω
2, CONT, FREQ, PERIOD, and
TEMP (thermocouple and thermistor). The 4-wire functions/operations include
Ω
4, TEMP
(4-wire RTD), RATIO, and CH AVG (ratio and channel average are covered in
).
With a 4-wire function/operation selected, channels are paired as follows:
• CH1 and CH11
• CH6 and CH16
2750-900-01.book Page 33 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...