3390 50 kHz Arbitrary Waveform / Function Generator Verification and Adjustment Manual Section 1: Verification procedures
3390-905-01Rev. B / May 2016
1-17
•
Scan INSTR:
Scan all of the connected equipment to see if all of the required instruments
are set.
•
Zero/Cal:
Power sensor calibration. Connect the power sensor to the POWER REF of
E4418B and then execute.
•
Begin:
Execute the manual calibration.
•
Address:
Show the address of each piece of calibration equipment.
Figure 7: ManualAdjustment.exe calibration address box
•
Secure AWG after adjusted:
Lock the manual calibration function after calibration is
complete.
•
Calibration completed.