Series 3700 System Switch/Multimeter Reference Manual
Section 7
:
Scanning
Document Number: 3700S-901-01 Rev. A / August 2007
7-7
Event blenders each have an associated event detector that can be accessed through script
control. Event blenders can only be accessed over the bus (no front panel control is available).
The following ICL commands provide additional information on available blenders:
trigger.blender[N].clear
(on page 9-216)
trigger.blender[N].orenable
(on page 9-216)
trigger.blender[N].overrun
(on page 9-216)
trigger.blender[N].stimulus[N]
(on page 9-217)
trigger.blender[N].wait
(on page 9-217)
Front panel scanning
After channels have been added to the scan list, pressing the
SCAN
key brings up the SCAN
ACTION MENU. If no scan list exists then pressing the
SCAN
key will briefly display "No Scan
List. Use INSERT to add selection." The SCAN ACTION MENU contains the following items:
EXECUTE:
Use this menu item to run scan list in the foreground.
LIST:
Use this menu item to see the current scan list steps. Turn the WHEEL to scroll
through the list.
CLEAR:
Use this menu item to clear the existing scan list.
RESET:
Use this menu item to reset the scan aspects of the unit. Scan aspects include
aspects such as scan count, clearing the scan list, and scan stimulus settings like scan
trigger arm.
Pressing the
INSERT
key will add the selected channel(s) or pattern to the existing scan list.
Pressing the
DELETE
key will remove the selected channel(s) or pattern from the existing scan
list. Only the first occurrence of the selected item will be removed. If Channel 3003 appears in
the list three times and Channel 3003 is selected when the
DELETE
key is pushed, the first step
using 3003 will be removed (the remaining two will stay in the list).
When removing channels, no check is performed to determine if a channel is associated with a
pattern. Continuing the previous example of 3003, if 'mypat1' was comprised of channels '3003,
3033, 3911 and 3922' when the remove request for 3003 is made, it will not touch or look at
'mypat1' if the pattern is in the list. To remove 'mypat1' from list, select the channel pattern
'mypat1' and press the when
DELETE
key.
Pressing the
STEP
key will single-step through a scan list.
Scan configuration
A scan is configured from the SCAN ATTR MENU. To access, while in an active scan list:
1. Press the
CONFIG
key.
2. Press the
SCAN
key. Modify any of the following menu items as desired:
ADD: This menu item specifies an additional list of channels and/or channel patterns to scan.
Related ICL command:
scan.add
(on page 9-157)