Section 10
:
Verification
Series 3700 System Switch/Multimeter Reference Manual
10-8
Document Number: 3700S-901-01 Rev. A / August 2007
5. Repeat steps 3 and 4 for each item in the table.
Figure 10-2: ACV verification
ACV verification data
Use the following values to verify the performance of the Series 3700. Actual values are
dependent on published specifications (see
Example reading limit calculation
(on page 10-3)).
Applied voltage
Range
Voltage
Lower limit
Upper limit
Verify ACV 100mV @ 20Hz
1.00E-01
1.00E-01
9.962000E-02
1.003800E-01
Verify ACV 100mV @ 1kHz
1.00E-01
1.00E-01
9.992000E-02
1.000800E-01
Verify ACV 100mV @ 50kHz
1.00E-01
1.00E-01
9.984000E-02
1.001600E-01
Verify ACV 100mV @ 100kHz
1.00E-01
1.00E-01
9.932000E-02
1.006800E-01
Verify ACV 1V @ 20Hz
1.00E+00
1.00E+00
9.992000E-01
1.00
Verify ACV 1V @ 1kHz
1.00E+00
1.00E+00
9.992000E-01
1.00
Verify ACV 1V @ 50kHz
1.00E+00
1.00E+00
9.984000E-01
1.00
Verify ACV 1V @ 100kHz
1.00E+00
1.00E+00
9.932000E-01
1.00
Verify ACV 10V @ 1kHz
1.00E+01
1.00E+01
9.00
1.01
Verify ACV 10V @ 50kHz
1.00E+01
1.00E+01
9.00
1.01
Verify ACV 10V @ 100kHz
1.00E+01
1.00E+01
9.00
1.01
Verify ACV 100V @ 1kHz
1.00E+02
1.00E+02
9.01
1.02
Verify ACV 100V @ 50kHz
1.00E+02
1.00E+02
9.01
1.02
Verify ACV 100V @ 100kHz
1.00E+02
1.00E+02
9.01
1.02
Verify ACV 300V @ 1kHz
3.00E+02
3.00E+02
2.02
3.02
Verify ACV 300V @ 50kHz
3.00E+02
3.00E+02
2.02
3.02