Series 3700 System Switch/Multimeter Reference Manual
Section 11
:
Calibration
Document Number: 3700S-901-01 Rev. A / August 2007
11-13
2. Send the following command:
SEND dmm.calibration.ac(7) print("done")
AC Cal Step 8: 100V @ 1kHz
1. Send the following command:
SEND dmm.range = 100
2. Source 100V @ 1kHz
3. Send the following command:
SEND dmm.calibration.ac(8) print("done")
AC Cal Step 9: 100V @ 50kHz
1. Source 100V @ 50kHz
2. Send the following command:
SEND dmm.calibration.ac(9) print("done")
AC Cal Step 10: 300V @ 1kHz
1. Send the following command:
SEND dmm.range = 300
2. Source 300V @ 1kHz
3. Send the following command:
SEND dmm.calibration.ac(10) print("done")