Series 3700 System Switch/Multimeter Reference Manual
Section 11
:
Calibration
Document Number: 3700S-901-01 Rev. A / August 2007
11-15
AC Cal Step 13: 10mA @ 1kHz
1. Send the following command:
SEND dmm.range = 10e-3
2. Source 10mA @ 1kHz
3. Send the following command:
SEND dmm.calibration.ac(13) print("done")
AC Cal Step 14: 100mA @ 1kHz
1. Send the following command:
SEND dmm.range = 100e-3
2. Source 100mA @ 1kHz
3. Send the following command:
SEND dmm.calibration.ac(14) print("done")
AC Cal Step 15: 1A @ 1kHz
1. Send the following command:
SEND dmm.range = 1
2. Source 1A @ 1kHz
3. Send the following command:
SEND dmm.calibration.ac(15) print("done")
AC Cal Step 16: 2A @ 1kHz
1. Send the following command:
SEND dmm.range = 2
2. Source 2A @ 1kHz
3. Send the following command:
SEND dmm.calibration.ac(16) print("done")