Section 3
:
Using the Front Panel
Series 3700 System Switch/Multimeter Reference Manual
3-14
Document Number: 3700S-901-01 Rev. A / August 2007
THERMO
Selecting the THERMO menu item opens the THERMO MENU. Items contained in this menu
are as follows:
REF JUNCTION: Allows selection of the Reference Junction to use. Available choices
are: SIMULATED, EXTERNAL, or INTERNAL. Related ICL command:
dmm.refjunction
(on page 9-109)
SIMREF: Specifies the simulated reference temperature for thermocouples. Related ICL
command:
dmm.simreftemperature
(on page 9-118)
THERMISTOR: Specifies the type of thermistor. Related ICL command:
dmm.thermistor
(on page 9-118)
COUPLE: Specifies the thermocouple type. Related ICL command:
dmm.thermocouple
(on page 9-119)
TRANSDUCER: Selects the transducer type (THERMOCOUPLE, THERMISTOR, 3RTD,
or 4RTD). Related ICL command:
dmm.transducer
(on page 9-121)
THREERTD: Specifies the type of 3-wire RTD. Related ICL command:
dmm.threertd
(on
page 9-119)
FOURRTD: Specifies the type of 4-wire RTD. Related ICL command:
dmm.fourrtd
(on
page 9-90)
USER: Specifies USER type of RTD (ALPHA, BETA, DELTA, or ZERO). Related ICL
commands:
dmm.rtdalpha
(on page 9-113),
dmm.rtdbeta
(on page 9-113),
dmm.rtddelta
(on page 9-114),
dmm.rtdzero
(on page 9-115)
THRESHOLD
Configures the threshold range. Related ICL command:
dmm.threshold
(on page 9-120)
UNITS
Configures the units for voltage and temperature measurements. Related ICL command:
dmm.units
(on page 9-121)