Section 13: LPT library function reference
Model 4200A-SCS Parameter Analyzer Reference Manual
13-160
4200A-901-01 Rev. C / February 2017
The pulse voltage low setting takes effect immediately during continuous pulse output. Otherwise, the
voltage low setting takes effect when the next trigger is initiated. The
pulse_trig
command is used
to trigger continuous or burst output.
The
pulse_vlow
,
pulse_vhigh
, and
pulse_dc_output
commands set the voltage value
output by the pulse channel when it is turned on (using
pulse_output
). If the output is already
enabled, these commands change the voltage level immediately, before the pulsing is started with a
pulse_trig
command.
Example
pulse_vlow(VPU1, 1, 0.5)
Sets the pulse voltage low value for channel 1 of the pulse card to 0.5 V.
Also see
(on page 13-140)
(on page 13-146)
(on page 13-152)
(on page 13-157)