Section 4: Multi-frequency capacitance-voltage unit
Model 4200A-SCS Parameter Analyzer Reference Manual
4-12
4200A-901-01 Rev. C / February 2017
3. Select
CVU Connection Compensation
.
Figure 88: CVU Connection Compensation dialog box
4. Select the
cable length
. You can select:
0 m:
Use if measurements are made at the terminals of the 4210-CVU (no cables).
1.5 m:
Use with the standard red SMA cables (part number CA-447A) that are supplied with
the 4210-CVU.
3 m:
Use with the red SMA cables (part number CA-446) that are supplied with the
4200-CVU-Prober-Kit. You can also use this setting if you are using a switching matrix.
Custom:
Cable length coefficients are measured by the user using the Measure Custom
Cable Length option under Advanced Compensation.
5. If you selected Custom cable length, select
Advanced Compensation
and select
Measure
Custom Cable Length
. Follow the on-screen instructions.
6. If you are using a switching matrix, close the matrix switches that connect the CVU to the open.
Refer to
(on page A-16).
7. In the Clarius CVU Connection Compensation dialog box, select
Measure Open
.
8. Follow the instructions.
9. Select
OK
.