Section 13: LPT library function reference
Model 4200A-SCS Parameter Analyzer Reference Manual
13-210
4200A-901-01 Rev. C / February 2017
Code
Status or error titles
-104
Unrecognized instrument or terminal not connected to matrix, argument #%d.
-105
No pathway assigned to argument #%d.
-106
Path %d previously allocated.
-107
Not enough pathways to complete connection.
-108
Argument #%d is not defined by configuration.
-109
Illegal test station: %d.
-110
A ground connection MUST be made.
-111
Instrument low connection MUST be made.
-113
There are no switching instruments in the system configuration.
-114
Illegal connection.
-115
Operation not allowed on a connected pin: %d.
-116
No physical bias path from %s to %s.
-117
Connection cannot be made because a required bus is in use.
-118
Cannot switch to high current mode while sources are active.
-119
Pin %d in use.
-120
Illegal connection between %s and GNDU.
-121
Too many calls were made to trigXX.
-122
Illegal value for parameter #%d.
-124
Sweep/Scan measure table overflow.
-126
Insufficient user RAM for dynamic allocation.
-129
Timer not enabled.
-137
Invalid value for modifier.
-138
Too many points specified in array.
-139
An error was encountered while accessing the file %s.
-140
%s unavailable while slaved to %s.
-141
Timestamp not available because no measurement was made.
-142
Cannot bind, instruments are incompatible.
-143
Cannot bind, services unavailable or in use.
-152
Function not supported by %s (%d).
-153
Instrument with ID %d is not in the current configuration.
-154
Unknown instrument name %s.
-155
Unknown instrument ID %i.
-158
VXI device in slot %d failed selftest (mfr ID: %04x, model number: %04x).
-159
VME device with logical address %d is either non-VXI or non-functional.
-160
Measurement cannot be performed because the source is not operational.
-161
Instrument in slot %d has non-functional dual-port RAM.
-164
VXI device in slot %d statically addressed at reserved address %d.
-165
Service not supported by %s (%d).
-166
Instrument with model code %08x is not recognized.
-167
Invalid instrument attribute %s.
-169
Instrument %s is not in the current configuration.
-190
Ill-formed connection.
-191
Mode conflict.
-192
Instrument sense connection MUST be made.
-200
Force value too big for highest range %g.