Model 4200A-SCS Parameter Analyzer Reference Manual
Section 4: Multi-frequency capacitance-voltage unit
4200A-901-01 Rev. C / February 2017
4-15
Generate load connection compensation data
Loads are reference re
sistors, typically 50 or 100 Ω or less, that must be resistive and constant over
the entire frequency range (1 kHz to 10 MHz). A load is connected to the output terminals using all
the cables, adaptors, probes and other hardware that will be in the test circuit.
To generate load correction data:
1. Make the connections to the 4210-CVU. See the
Test connections for a switch matrix
(on page 4-
2. In Clarius, select
Tools
. The Clarius Tools dialog box opens.
Figure 92: Clarius Tools dialog box
3. Select
CVU Connection Compensation
.
Figure 93: CVU Connection Compensation dialog box
4. Select the
cable length
. You can select:
0 m:
Use if measurements are made at the terminals of the 4210-CVU (no cables).
1.5 m:
Use with the standard red SMA cables (part number CA-447A) that are supplied with
the 4210-CVU.
3 m:
Use with the red SMA cables (part number CA-446) that are supplied with the
4200-CVU-Prober-Kit. You can also use this setting if you are using a switching matrix.
Custom:
Cable length coefficients are measured by the user using the Measure Custom
Cable Length option under Advanced Compensation.