Model 4200A-SCS Parameter Analyzer Reference Manual
Appendix A: Using switch matrices
4200A-901-01 Rev. C / February 2017
A-29
Figure 543: Assign switch cards to slots
Step 7. Set matrix card properties
The matrix card properties set the connections:
•
Between the measurement instrumentation and the matrix card
•
Between the matrix card and the test system (prober or test fixture)
The number of pins defined in the properties for a probe station or test fixture determines the pins
that are available to assign to a switch matrix card column. Make sure the number of pins assigned
is appropriate for your system. Refer to
page A-26) for additional information.
To set matrix card properties:
1. In the Configuration Navigator, expand the switching matrix.
2. Select the card. The properties are displayed. Each row and column has a list to set the card
properties. If the row-column connection scheme is selected, instruments are assigned to the
rows and the test fixture pins or probe pins are assigned to the columns. If the instrument card
connection scheme is selected, both instrumentation and test fixture/probe pins are assigned to
columns.
The following figure shows the KI 7XXX matrix-card
Properties
tab configuration that is required
to support the physical connection configuration that is shown in
(on page A-13).
3. Select from the lists to connect the rows and columns to instrument terminals and prober or test
fixture pins. Card properties must match the actual physical connections to the matrix card.
In the figure below, the lists labeled
A
to
H
correspond to the eight rows of all Keithley
Instruments Series 700 Switching System-compatible matrix cards. The lists labeled
1
to
12
correspond to the 12 columns of Series 700 Switching System-compatible matrix cards.
4. Select
Validate Configuration
.
Prober or test-fixture pins are always connected to matrix-card columns.