Section 4: Multi-frequency capacitance-voltage unit
Model 4200A-SCS Parameter Analyzer Reference Manual
4-16
4200A-901-01 Rev. C / February 2017
5. If you selected Custom cable length, select
Advanced Compensation
and select
Measure
Custom Cable Length
. Follow the on-screen instructions.
6. If you are using a switching matrix, close the matrix switches that connect the CVU to the open.
Refer to
(on page A-16).
7. If it is not open, select
Advanced Compensation
.
8. Next to Measure Load, enter the value of the load in ohms.
9. Select
Measure Load
.
10. Follow the instructions.
11. Select
OK
.
Compensation data
You can view the compensation data. Clarius lists R and jX compensation values for every test
frequency and measurement range for open, short, and load.
To view the data generated by connection compensation:
1. In Clarius, select
Tools
.
2. Select
CVU Connection Compensation
.
3. Select
Advanced Compensation
.
4. Select the data you would like to display:
Open
,
Short
, or
Load
.
5. Select
View Compensation Data
.
6. Select the
HI
tab to review the high side values.
7. Select the
LO
tab to review the low side values.
Figure 94: Open compensation values example